• Gossen Metrawatt

SECULIFE ST BASE25

Vielen Dank für Ihr Interesse! Wir werden uns umgehend um Ihre Anfrage kümmern.

Data protection

We attach great importance to data protection!
Read more here in our Privacy Statement

Item number: M694B

Test device with antimicrobial effectiveness, 25 A test current and standard-compliant test sequences for testing the electrical safety of medical devices (DGUV regulation 3, MPBetreibV)

The SECULIFE ST BASE 25 supports all required measurement and test functions for testing the effectiveness of protective measures in medical devices and systems - both with predefined and programmable test sequences (VDE 0701-0702, IEC/EN 62353, VDE 0751). The test current of up to 25A allows the device to be used for type testing. The housing is equipped with an antimicrobial effect which enables the device to be used even in hygienically sensitive areas.
  • Test device with antimicrobial effectiveness, 25 A test current and standard-compliant test sequences for testing the electrical safety of medical devices (DGUV regulation 3, MPBetreibV)
  • Test device with antimicrobial effectiveness, 25 A test current and standard-compliant test sequences for testing the electrical safety of medical devices (DGUV regulation 3, MPBetreibV)
  • Test device with antimicrobial effectiveness, 25 A test current and standard-compliant test sequences for testing the electrical safety of medical devices (DGUV regulation 3, MPBetreibV)

The SECULIFE ST BASE 25 was specially developed for all demanding testing tasks in hygienically sensitive areas of medical technology, the pharmaceutical industry and the food industry. New optimally adapted measuring cycles ensure faster and more efficient measuring and testing. Programmable test sequences and pre-defined sequences that comply with the required standards and regulations allow a simple and intuitive workflow. With the SECULIFE ST BASE 25 it is possible to carry out tests without special knowledge and this with the quality of a leading test instrument concept for decades.

Product Highlights

  • Antimicrobial effectiveness of front and housing of the tester against bacteria, viruses and fungi
  • Pre-configured and self-programmable test sequences for fast testing of electrical devices and medical devices
  • Comprehensive data management and storage concept
  • Automatic test sequence and individual measurements for up to 50,000 data records
  • High resolution and brilliant 4,3"" TFT colour display
  • Create your own test sequences with the help of "SequenceDesigner" (free download in myGMC)

Your advantages

  • User guidance through pre-programmed test sequences according to standard with automatic evaluation - thus excellently suited for use by instructed persons
  • Quick access to measurement and test functions via rotary switches, direct selection keys and SoftKeys
  • The compact, shock-proof housing with integrated rubber protection is particularly suitable for mobile use on construction sites and in industrial environments
  • The unique multiple measurement allows convenient documentation of several measuring points
  • Extensive setting options for language, keyboard, character set, date, time as well as country-specific standard versions, also suitable for international use
  • For maximum IZYTRONIQ compatibility, a DUT can be classified in the location tree in addition to the customer relationship (property, building, level, room). Additional fields such as department and cost center are available.
  • Multiprint - Printout of all protocols of a test object at the touch of a button (on thermal printer Z721S)
  • Create user-defined test sequences using the IZYTRONIQ software

Applications

  • Safe, comprehensive and efficient testing of electrical devices according to manufacturer specifications through programmable test sequences of any standards.
  • Testing of electrical medical devices according to IEC/EN 62353 (VDE 0751).
  • Testing of electrical devices according to DIN VDE 0701-0702 including (extension) cables and PRCDs.
  • Technical characteristics
    • Protective conductor resistance
    • Insulation resistance measurement
    • Leakage current measurements: Protective conductor current, device leakage current, touch current, patient leakage current,
    • Leakage current from application part, as replacement (alternative) methods, direct method, differential method
    • Voltage measurement: Contact voltage, SELV, PELV voltage PRCD Tripping time
    • Function test (U,I,P,S,LF,f)
    • Line test: continuity, short circuit, polarity (wire exchange)
    • Temperature measurement (for Pt100/Pt1000 sensor)
  • Particularities
    • Equipping the housing and front foil with an anti-microbial effect to prevent colonisation by viruses, bacteria or fungi Preprogrammed
    • Procedures (according to VDE 0701-0702) for testing PRCDs of type PRCD standard, SPE-PRCD, PRCD-S and PRCD-K
    • Remote control possible
    • Additional database elements property, building, level and room in order to be able to structure more extensive databases better as well as additional fields department + cost center
    • Multiprint - Printout of all protocols of a test object with 1 keystroke (on thermal printer Z721S or on USB stick)
    • Create user-defined protocol templates (for thermal printer Z721S or USB stick) using "ReportDesigner" (free download in myGMC)
    • RFID transponder (Z751R,S,T) read and write (with SCANBASE RFID Z751E)
    • Data export ETC file (XML) to USB stick
    • Data import of all test object master data into test device from ETC or USB stick (ETC file)
    • 25A Test current

Accessories

Contact

Do you have any questions on the product or do you need technical information or details on the current delivery period? Our sales backoffice will be pleased to assist you.

Sales International:
Telephone: +49 911 8602-0
Send message

myGMC

Under myGMC you can register your product and receive up-to-date information on firmware updates, modifications of standards and many more topics.

Dialogue, Information and Actions: